NIOSH Manual of Analytical Methods (1994)
National Institute for Occupational Safety and Health
Lead by Field Portable XRF (7702)
2003076NIOSH Manual of Analytical Methods — Lead by Field Portable XRF (7702)1994National Institute for Occupational Safety and Health

LEAD BY FIELD PORTABLE XRF Pb

MW: 207.19 (Pb) 223.19 (PbO)

METHOD: 7702, Issue 1

CAS: 7439-92-1 (Pb) 1317-3608 (PbO)

RTECS: OF7525000 (Pb) OG1750000 (PbO)

EVALUATION: FULL

OSHA: 0.05 mg/m3 NIOSH: <0.1 mg/m3; blood lead 60 µg/100 g ACGIH: 0.05 mg/m3; BEI blood 30 µg/ 100 mL

PROPERTIES:

7702

Issue 1: 15 January 1998

soft metal; d 11.3 g/mL @ 20 C; MP 327.5 C; BP 1740 C; valences +2, +4 in salts

SYNONYMS: elemental lead and lead compounds except alkyl lead

SAMPLING SAMPLER:

MEASUREMENT

FILTER (0.8-µm, 37-mm, mixed cellulose ester membrane)

TECHNIQUE:

FLOW RATE: 1 to 4 L/min

X-RAY FLUORESCENCE (XRF), PORTABLE, L-SHELL EXCITATION (e.g., 109 Cd source) NOTE: Performance parameters are based upon research conducted with the NITON® 700 XRF [1].

3

VOL-MIN: -MAX:

570 L @ 30.0 µg/m [1] 1900 L @ 9.0 µg/m3

SHIPMENT:

routine

ANALYTE:

lead

SAMPLE STABILITY:

stable

CALIBRATION:

lead thin-film standards (Micromatter Co., or equivalent); internal instrument calibration

BLANKS:

2 to 10 field blanks per set RANGE:

17 to 1500 µg of Pb per sample [1]

ACCURACY ESTIMATED LOD: 6 µg of Pb per sample [1] RANGE STUDIED: 0.1 to 1514.6 µg/m3 (as Pb) (based upon lead mass loadings) BIAS: PRECISION ( ACCURACY:

0.069 [1] rT

):

0.054 @ 10.3 to 612 µg Pb per sample ±16.4%

APPLICABILITY: This method was evaluated for air samples on filters only. The working range of this method is 0.017 mg/m3 to 1.5 mg/m3. This is a field portable analytical method, particularly useful for the analysis of initial exposure assessment samples, or for applications where laboratory analysis is impractical. Additionally, the method is non-destructive; samples analyzed in the field can later be analyzed in a laboratory. The method is applicable to all elemental lead forms, including lead fume, and all other aerosols containing lead. INTERFERENCES: The presence of bromine will cause XRF readings for lead to be elevated, resulting in a positive bias error. Other interferences may exist in other XRF instruments. OTHER METHODS: Laboratory-based methods include atomic spectrophotometric methods following hot plate acid digestion: NIOSH methods 7082 (flame atomic absorption spectrophotometry) [2], 7105 (graphite furnace atomic absorption spectrophotometry) [3], and 7300 (inductively coupled plasma atomic emission spectrophotometry) [4]. A field-portable analytical method for lead air filter samples using ultrasound/ASV has been developed, NIOSH Method 7701[5]. A field-portable screening method by spot test kit has been developed, NIOSH Method 7700 [6].

NIOSH Manual of Analytical Methods (NMAM), Fourth Edition LEAD BY FIELD PORTABLE XRF: METHOD 7702, Issue 1, dated 15 January 1998 - Page 2 of 4 REAGENTS:

EQUIPMENT:

1. None

1. Sampler: Mixed cellulose ester filter, 0.8-µm pore size, 37-mm diameter, with cellulose backup pad, in a closed-faced cassette filter holder. 2. Personal sampling pump, 1 to 4 L/min, with flexible connecting tubing. 3. Field portable, L-shell X-Ray Fluorescence (XRF) instrument with a Cadmium-109 source. 4. Filter sleeve: thin cardboard with 37-mm dia. cut out, and covered with a light adhesive between two pieces of acetate (Mylar™) (NITON, Bedford, MA, or equivalent). NOTE: Material must be transparent to X-ray. 5. Filter test platform to hold the filter (specific to instrument). 6. Forceps 7. Thin film standard reference materials from 15 µg/cm2 to 150 µg/cm2 (Micromatter Co., Deer Harbor, WA), or equivalent [7,8].

SPECIAL PRECAUTIONS: None

SAMPLING: 1. 2.

Calibrate each sampling pump with a representative sampler in line. Sample at an accurately known flow rate (1 to 4 L/min) for a total sample size of approximately 1000 L. Do not exceed a filter loading of 2 mg total dust.

SAMPLE PREPARATION: 3.

4. 5.

With forceps, transfer the MCE filter without the backup pad to a filter sleeve. The sleeve material must be transparent to X-rays (see EQUIPMENT, Item 4). NOTE: Take special care when removing the filter from the backup pad to avoid loss of leadcontaining dust. Place the filter into 37-mm opening and seal with Mylar™ film to prevent losses and allow undisturbed analysis of the filter. Place the sealed filter onto the filter test platform of the instrument for analysis. NOTE: The NITON® 700 Series XRF has a filter testplatform that allows for three readings with no substrate effect.

CALIBRATION AND QUALITY CONTROL: 6.

Start XRF and allow a 30-minute warm-up period. The instrument will conduct an internal selfcalibration. 7. Using thin film standards [8], verify the internal calibration to within ± 5% of the calibration standard. 2 Use a minimum of three standards at concentrations of 15 µg/cm , 150 µg/cm2, and one standard concentration between these two values. 8. Restart the instrument as needed to assure instrument accuracy prior to sample analysis. NOTE: When the thin film standard measurements are not within the specified parameters, the instrument may need to be recalibrated at the factory. 9. Analyze one thin film standard every 2 hours to check for instrument drift. 10. Repeat step 7 when all analyses are completed as a post-calibration check. MEASUREMENT: NIOSH Manual of Analytical Methods (NMAM), Fourth Edition LEAD BY FIELD PORTABLE XRF: METHOD 7702, Issue 1, dated 15 January 1998 - Page 3 of 4

11. Set instrument parameters and analyze filter samples as specified by the manufacturer. The following measurement technique is based upon the NITON® 700 XRF. a. Analyze the middle of the sample filter first (see Figure 1, M). b. Allow the instrument to take a one source-minute reading (This may take longer than one real-time minute, depending upon the source strength). A one source-minute reading will assure the accurate L-shell reading necessary for the analysis of lead air filter samples. c. Analyze the filter sample at the top of the filter for one source minute (see Figure 1, T). d. Analyze the filter sample at the bottom of the filter for one source minute (see Figure 1, B). 2 e. The instrument software uses an algorithm that converts the three readings in µg/cm to an analytical result in µg of lead per sample. This result will be displayed following the third filter reading [1]. f. Analyze one standard every 2 hours (step 8). g. Repeat three-reading calibration check following completion of analyses (step 8). CALCULATIONS: 3 12. Using the measured lead concentration, W (µg), calculate the concentration, C (mg/m ), of lead in the air volume sampled, V (L):

C

NOTE: µg/L

W , mg/m 3 V

mg/m3

EVALUATION OF METHOD: This method was validated on field samples [1] by collecting lead particulate samples from bridge lead abatement projects. Airborne concentrations of lead within the containment of a sand blasting bridge 3 lead abatement project ranged from 1 to 10 mg/m . Area samples were collected for periods of time ranging from 15 seconds to 2 hours. This sampling protocol yielded 61 filter samples with lead loadings ranging between 0.1 to 1514.6 µg of lead per sample. Four personal samples were collected from a hand-scraping bridge lead abatement project for a total sample size of 65. The samples were first analyzed using a non-destructive, field portable XRF method. Samples subsequently were subjected to confirmatory analysis by the laboratory based NIOSH method 7105, Lead by GFAAS [3]. The method was statistically evaluated according to the NIOSH Guidelines for Air Sampling and Analytical Method Development and Evaluation [9]. The overall precision (rT) of the XRF method was calculated at 0.054 with a 95% confidence interval (CI) of 0.035 to 0.073, and the bias was 0.069 with a 95% CI of 0.006 to 1.515. The XRF method accuracy was determined to be ± 16%; however, at the upper 90% CI, the accuracy is ± 27%. Since the confidence interval includes the ± 25%, meeting the NIOSH accuracy criteria of ± 25% is inconclusive. However, the samples used to evaluate this method were field samples. Laboratory prepared aerosol samples would be expected to give better precision. Additionally, the XRF method is non-destructive; samples analyzed in the field can subsequently be analyzed in a laboratory using a method with greater accuracy, as needed. The filter sleeve used with the NITON® 700 Series XRF used a Mylar film to cover and seal the 37-mm filter. The lead particulate on the surface of the filter came into contact with the Mylar™ film. Both the Mylar™ film and the filter were digested with nitric acid and hydrogen peroxide as is specified in NIOSH Method 7105 [3].

REFERENCES: [1]

[2]

Morley JC [1997]. Evaluation of a portable x-ray fluorescence instrument for the determination of lead in workplace air samples [Thesis]. Cincinnati, OH: University of Cincinnati, Department of Environmental Health, College of Medicine. NIOSH [1994]. Lead by FAAS: Method 7082. In: Eller PM, Cassinelli ME, eds. NIOSH Manual of Analytical Methods (NMAM), 4th ed. Cincinnati, OH: National Institute for Occupational Safety and NIOSH Manual of Analytical Methods (NMAM), Fourth Edition LEAD BY FIELD PORTABLE XRF: METHOD 7702, Issue 1, dated 15 January 1998 - Page 4 of 4

[3]

[4]

[5]

[6]

[7]

[8] [9]

Health, DHHS (NIOSH) Publication No. 94-113. NIOSH [1994]. Lead by GFAAS: Method 7105. In: Eller PM, Cassinelli ME, eds. NIOSH Manual of Analytical Methods (NMAM), 4th ed. Cincinnati, OH: National Institute for Occupational Safety and Health, DHHS (NIOSH) Publication No. 94-113. NIOSH [1994]. Elements: Method 7300. In: Eller PM, Cassinelli ME, eds. NIOSH Manual of Analytical Methods (NMAM), 4th ed. Cincinnati, OH: National Institute for Occupational Safety and Health, DHHS (NIOSH) Publication No. 94-113. NIOSH [1994]. Lead by ultrasound/ASV: Method 7701 (supplement issued 1/15/98). In: Eller PM, Cassinelli ME, eds. NIOSH Manual of Analytical Methods (NMAM),th4ed. Cincinnati, OH: National Institute for Occupational Safety and Health, DHHS (NIOSH) Publication No. 98-119. NIOSH [1994]. Lead in air by chemical spot test: Method 7700 (supplement issued 5/15/96). In: Eller PM, Cassinelli ME, eds. NIOSH Manual of Analytical Methods (NMAM),th4ed. Cincinnati, OH: National Institute for Occupational Safety and Health, DHHS (NIOSH) Publication No. 96-135. National Institute of Standards and Technology, Standard Reference Material Program, Standard Reference Material 2579, Lead paint film on Mylar sheet for portable X-ray fluorescence analyzers, Gaithersburg, MD 20899. Micromatter Co., XRF Calibration Standards, P.O. Box 123, Deer Harbor, Washington, 98243, Phone (360) 3 76-4007. NIOSH [1995]. Kennedy, E.R., Fishbach, T.J., Song, R., Eller, P.M., Shulman, S.S.. Guidelines for Air Sampling and Analytical Method Development and Evaluation. Cincinnati, Ohio: National Institute for Occupational Safety and Health, DHHS (NIOSH) Publication No. 95-117.

METHOD WRITTEN BY: J. Clinton Morley, MS, NIOSH/DSHEFS

,

Figure 1: Analysis of a 37-mm filter (XRF windows identified as M, T,

using XRF and B are 2 cm x 1 cm)

NIOSH Manual of Analytical Methods (NMAM), Fourth Edition