P-Toluenesulfonic Acid (5043)

From Wikisource
Jump to navigation Jump to search
NIOSH Manual of Analytical Methods (1994)
National Institute for Occupational Safety and Health
P-Toluenesulfonic Acid (5043)
2002986NIOSH Manual of Analytical Methods — P-Toluenesulfonic Acid (5043)1994National Institute for Occupational Safety and Health

p-TOLUENESULFONIC ACID

C7H8O3S

MW: 172.20

METHOD: 5043, Issue 1

5043

CAS: 104-15-4

RTECS: XT6300000

EVALUATION: PARTIAL

OSHA : no PELs NIOSH: no RELs ACGIH: no TLVs (1 ppm = 7.03 mg/m3 @ NTP)

PROPERTIES:

Issue 1: 15 January 1998

solid; MP 106 to 107 C when anhydrous; d 1.20 g/mL @ 20 C; BP 140 C @ 20 mm Hg; VP not significant @ 25 C

NAMES & SYNONYMS: 4-methylbenzenesulfonic acid; p-methylphenylsulfonic acid; tosic acid

SAMPLING

MEASUREMENT

SAMPLER:

FILTER (13-mm glass fiber)

FLOW RATE:

1 to 3 L/min

VOL-MIN: -MAX:

10 L @ 1000 L

SHIPMENT:

routine

SAMPLE STABILITY:

at least 29 days @ 25 C [1]

BLANKS:

2 to 10 field blanks per set

EXTRACTION:

2 mL 2% isopropanol in water v/v; ultrasonic bath, 10 min

100 µL

COLUMN:

3.9-mm ID x 15 cm stainless steel packed with 10-µm µ-Bondapak C18

DETECTOR:

UV @ 222 nm

CALIBRATION:

standard solutions of ptoluenesulfonic acid in isopropanol: water

RANGE:

0.27 to 120 µg/sample [1]

not determined

ESTIMATED LOD:

0.08 µg/sample [1]

not determined

PRECISION ( r):

0.046 @ 3 to 15 µg/sample [1]

BIAS:

not determined

ACCURACY:

p-toluenesulfonic acid

21% acetonitrile/79% water containing 0.005 M Pic® A v/v; 1 mL/min

not studied

):

ANALYTE:

MOBILE PHASE:

RANGE STUDIED:

rT

HPLC, UV DETECTOR

INJECTION VOLUME:

ACCURACY

OVERALL PRECISION (

TECHNIQUE:

APPLICABILITY:p-Toluenesulfonic acid at 0.0027 to 1.2 mg/m3 can be determined in a 100-L air sample by this method. This method has been used to analyze samples collected at a foundry & machining company where p-toluenesulfonic acid was a catalyst in a process to produce a resin from resorcinol and furfuryl alcohol [2].

INTERFERENCES: None identified.

OTHER METHODS: Chromatographic methods for determination of p-toluenesulfonic acid in solution have been published [3-7]. None has been published for air analysis. An alternative air sampler for the present method is a midget impinger containing isopropanol [1,2].

NIOSH Manual of Analytical Methods (NMAM), Fourth Edition p-TOLUENESULFONIC ACID: METHOD 5043, Issue 1, dated 15 January 1998 - Page 2 of 4 REAGENTS:

EQUIPMENT:

1. 2. 3. 4. 5.

1. Sampler: 13-mm glass fiber filter in 2-piece filter holder (Swinnex, Millipore Corp., or equivalent). 2. Personal sampling pump, 1 to 3 L/min, with flexible connecting tubing. 3. HPLC, UV detector, 222 nm, integrator, and column (page 5043-1). 4. Vials, 4-mL, with PTFE-lined caps. 5. Graduated cylinder, 1-L, readable to 10 mL. 6. Volumetric flasks, 25- and 10-mL. 7. Syringes, 10-mL, 500-µL, and 100-µL. 8. Ultrasonic water bath. 9. Film, plastic, water resistant. 10. Syringe filters, 3-mm PTFE membranes, 0.45µm pore size, in polypropylene housing. 11. Forceps.

p-Toluenesulfonic acid,* >98% pure. Water, distilled. Acetonitrile,* chromatographic quality. Isopropanol, chromatographic quality. PIC®A, “paired-ion chromatography,” low UV reagent (Waters Corp., or equivalent). PIC®A contains water, tetrabutylammonium hydrogen sulfate, and phosphoric acid. Add entire vial of PIC®A to 1 L of distilled water to make 0.005M solution. 6. Extraction solution: 2% Isopropanol/98% water (v/v). Add 20 mL of isopropanol to distilled water to make 1 L of solution. 7. p-Toluenesulfonic acid calibration stock solution, 4.0 mg/mL. Dissolve 200 mg ptoluenesulfonic acid in extraction solution to make 50 mL of solution.

  • See SPECIAL PRECAUTIONS

SPECIAL PRECAUTIONS: p-Toluenesulfonicacid is highly irritating to the skin and mucous membranes; also, it can cause a potentially explosive reaction of acetic anhydride with water. Acetonitrile is toxic and is a fire hazard (flash point = 12.8 C). Wear protective clothing and work in a well ventilated hood.

SAMPLING: 1. Calibrate each personal sampling pump with a representative sampler in line. 2. Attach the sampler to the personal sampling pump with flexible tubing. 3. Sample at an accurately known flow rate between 1 and 3 L/min for a total sample size of 10 to 1000 L. NOTE: Limit the maximum loading of particulate matter on the filter to approximately 0.5 mg. 4. Seal ends of sampler with plastic film. Ship to laboratory at room temperature.

SAMPLE PREPARATION: 5. 6. 7. 8.

Using the forceps, transfer the 13-mm glass fiber filter to a 4-mL vial. Add 2 mL of the extraction solution to the vial and cap securely. Place the sample vial into an ultrasonic bath and agitate for 10 min. Filter the sample solution through a PTFE membrane syringe filter.

CALIBRATION AND QUALITY CONTROL: 9. Calibrate daily with at least six working standards over the range of interest: 0.04 to 60 µg/mL. a. Prepare working standards over the range of interest. Add a known amount of the calibration stock solution to an aliquot of the extraction solution. b. Analyze together with samples and blanks (steps 12 and 13). c. Prepare calibration graph (peak area or height vs. µg of analyte). 10. Determine recovery (R) at least once for each lot of glass fiber filters in the calibration range (step 9). NIOSH Manual of Analytical Methods (NMAM), Fourth Edition p-TOLUENESULFONIC ACID: METHOD 5043, Issue 1, dated 15 January 1998 - Page 3 of 4 Prepare three filters at each of five concentration levels plus three media blanks. a. Place 13-mm glass fiber filters into 4-mL vials. b. With a microliter syringe, fortify each filter with a known amount of the calibration stock solution. c. Allow the uncapped vials to stand overnight at room temperature. d. Prepare and analyze with working standards (steps 5 through 8, and steps 12 and 13). e. Prepare graph of R vs. µg of analyte recovered. 11. Analyze three quality control blind spikes and three analyst spikes to ensure that the calibration and recovery graphs are in control.

MEASUREMENT: 12. Set high performance liquid chromatograph to manufacturer’s recommendations and to conditions given on page 5043-1. Inject 100-µL aliquot manually or with autosampler. NOTE: If peak area is above the range of the working standards, dilute with desorbing solution, reanalyze, and apply appropriate dilution factor in calculations. 13. Measure peak area or height forp-toluenesulfonic acid.

CALCULATIONS: 14. Determine the mass, µg (corrected for R), of analyte found on the filter (W) and the average media blank (B). 15. Calculate the concentration, C, ofp-toluenesulfonic acid in the air volume sampled, V (L):

C

W

B V

, mg/m 3

EVALUATION OF METHOD: Average recoveries ofp-toluenesulfonic acid after fortification of 13-mmglass fiber filters with 3-, 6-, 10-, and 15-µg quantities of the compound were 1.02, 0.96, 1.04 and 0.96, respectively; precision r() was 0.046 (23 samples, pooled). After 29 days of storage at room temperature, the average recovery of 3-µg quantities of p-toluenesulfonic acid from glass fiber filters was 1.03; rSwas 0.023 (5 samples). The purpose of the 2% isopropanol in samples and standards was to prevent possible deterioration of p-toluenesulfonic acid by bacteria during storage.

ALTERNATIVE METHOD: As an alternative to the glass fiber filters, air samples ofp-toluenesulfonic acid canbe collected in midget impingers containing isopropanol at 1 L/min. After sampling, the impinger solution is transferred to a 20-mL glass vial and transported to the laboratory. The vial is placed onto a heating plate maintained at 80C, and the isopropanol is evaporated to dryness with a gentle stream of nitrogen (total evaporation time is about 1.5 hours). A 2-mL aliquot of 2% isopropanol in water is added to the vial, and the vial is placed into an ultrasonic bath for 30 seconds. Then the vial is tilted in order to wet the inside wall of the vial. Solution is filtered through a PTFE membrane syringe filter and is ready for analysis according to the method described above.

EVALUATION OF ALTERNATIVE METHOD: Isopropanol was found to evaporate from the midget impinger during air sampling. Air at room temperature was drawn at 1 L/min through an impinger containing 20 mL of isopropanol. After 1 hour of pump operation, NIOSH Manual of Analytical Methods (NMAM), Fourth Edition p-TOLUENESULFONIC ACID: METHOD 5043, Issue 1, dated 15 January 1998 - Page 4 of 4 6 mL of the isopropanol had evaporated, and the remaining 14 mL was cold because of evaporation. Separate recovery experiments were performed. The average recovery of 3-µg quantities ofptoluenesulfonic acid from 20-mL volumes of fortified isopropanol was 1.04 (Sr = 0.054 for 6 samples); the average recovery of 6-µg quantities ofp-toluenesulfonic acid from20-mL volumes of fortified isopropanol was 1.01 (Sr = 0.028 for 6 samples). After 29 days of storage at room temperature, the average recovery of 3-µg quantities of p-toluenesulfonic acid from 20-mL volumes of isopropanol was 0.95 (S r = 0.050 for 6 samples). This alternative method and the method with glass fiber filters both have the same LOD and LOQ, because final sample solution volumes in the two methods are the same.

REFERENCES: [1] [2] [3]

[4]

[5] [6] [7]

Tucker SP [1997]. Backup data report for method 5043,p-Toluenesulfonic Acid. National Institute for Occupational Safety and Health, DPSE/NIOSH, (Unpublished report). Tucker SP [1996]. Analytical report for Sequence #8567, DPSE/MRSB. Cincinnati, OH: National Institute for Occupational Safety and Health. ( Unpublished report). Lange FT, Wenz M, Brauch H [1995]. Trace-level determination of aromatic sulfonates in water by online ion-pair extraction/ion-pair chromatography and their behavior in the aquatic environment. J High Resolut Chromatogr18:243-252. Altenbach B, Giger W [1995]. Determination of benzene- and naphthalene sulfonates in wastewater by solid-phase extraction with graphitized carbon black and ion-pair liquid chromatography with UV detection. Anal Chem67:2325-2333. Bastian B, Knepper TP, Hoffmann P, Ortner HM [1994]. Fresenius’ J Anal Chem 348: 674-679. Brouwer ER, Slobodnik J, Lingeman H, Brinkman UAT [1992]. Analusis 20: 121-126. Schullerer S, Brauch HJ, Frimmel FH [1990]. Vom Wasser75:83-97.

METHOD WRITTEN BY: Samuel P. Tucker, Ph.D., NIOSH/DPSE

NIOSH Manual of Analytical Methods (NMAM), Fourth Edition