interviews during the onsite visit, these freezers were identified as being used for CLIA activities. The submission did not include a response or data related to the freezers identified as "not used in patient testing."

To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit," the means by which a "tracer audit" would be documented, and whether the results of a "tracer audit" would be the information reviewed by the QMPI Program.

The laboratory failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information set forth above and in our January 25, 2016 letter.

Finding #2
The submission references "Ex. I, Tabs 2-6." We located these tabs, but found no documentation in Tabs 2, 5, and 6.

Although the laboratory's submitted protocol indicates that the laboratory was required to check manufacturer package inserts prior to use (see Ex. A, Tab 31, §8.1.2), which was also stated in the laboratory's written procedure available prior to the onsite survey, the laboratory provided no documentation showing that this protocol has been effectuated and no indication that package inserts for new lot numbers have been checked. In addition, the submission states that "the lab has conducted training on those procedures;" however, the training documents submitted in Ex. A, Tab 32 do not include any training specific to the cited deficiency. We were unable to verify that training occurred as stated.

In Ex. I, Tab 1, the laboratory provided lists of patient specimen accession numbers for which the laboratory issued corrected reports; however, the laboratory did not provide copies of the corrected reports.

The laboratory failed to address and provide acceptable evidence of correction consisting of: what measure has been put in place or what systemic changes have been made to ensure the deficient practice does not recur; and, how the corrective action(s) is being monitored to ensure the deficient practice does not recur.

D5421
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

Finding #1 The submission references "Ex. A, Tab 9, §7.2.4.4 and Ex. B, Tabs 46-50, 52-56, 57-61." We found no such references contained in the materials provided to CMS.