The submission (Ex. E) includes a list of patient specimen accession numbers for which the laboratory "identified reports," but does not specify or include documentation to indicate whether corrected reports were generated and issued.

The laboratory failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information set forth above and in our January 25, 2016 letter.

Finding #3
The laboratory's review of the QC found in Ex. E, Tab 1 did not specify what period of QC the laboratory reviewed. The quality assessment (QA) documentation related to QC data from the survey in July 2014, October 2014, and February through June 2015. The laboratory's investigation did not indicate that the review had identified the cited issues related to QC.

To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit," the means by which a "tracer audit" would be documented, and whether the results of a "tracer audit" would be the information reviewed by the QMPI Program.

The laboratory failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information set forth above and in our January 25, 2016 letter.

Finding #4
Although the laboratory's submitted protocol included information about quarterly audits and agenda items for the laboratory's QMPI Program, the submission does not provide specific actions to take when problems are identified, specifically related to clotted specimens. The laboratory states in Ex. L, Tab 35 that "[p]atient specimens that did not meet the lab's acceptance criteria were rejected. . ." However, the issue was not that the clotted specimens were rejected, rather that the laboratory's QA program did not identify the high number of clotted specimens over a period of three quarters. The submission does not include any documentation showing that the issue was identified and corrected. We note that Ex. A, Tab 12, §7.2.1.6 indicates that the QMPI Program agenda included specimen rejection rate partitioned by rejection criteria, but does not require that any action be taken based on this agenda item. In addition, we note that in Ex. A, Tab 12, §7.1, the laboratory provided the frequency of the QMPI Program meetings, but does not indicate any action should be taken in the event it is determined that the quality metrics were not met.

To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit,” the means by

23