California. The laboratory relied on its written policies and procedures to ensure that appropriately licensed testing personnel were performing and reporting patient test results. However, if the laboratory's policies and procedures are inconsistent with state requirements, we question whether this mechanism would prevent the deficient practice from recurring, and question the laboratory's document control system to ensure accurately written policies and procedures.

The laboratory failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information set forth above and in our January 25, 2016 letter.

D6171
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

Ex. L, Tab 3 of the submission states: "TP14 has been retrained to ensure that TP14 only performs activities within the scope of TP14's job description as a clinical laboratory associate, under the supervision of testing personnel. Because TP14 is not testing personnel, the high complexity personnel educational requirements do not apply." The CLIA job responsibilities in the job description provided at Ex. L, Tab 3 does include CLIA regulatory responsibilities for testing personnel, and as such TP14 was required to meet the educational requirements. The laboratory's submission does not provide any additional documentation that establishes that TP14 was qualified to perform high complexity testing. TP14 remains unqualified to perform high complexity testing.

The laboratory failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information set forth above and in our January 25, 2016 letter.

D6178
The laboratory's allegation of compliance is not credible and evidence of correction is not acceptable.

Ex. F contains lists of patient specimen accession numbers for which the laboratory intended to issue corrected test reports. The laboratory provided no documentation to indicate corrected reports were generated and issued.

To ensure the deficient practice does not recur, the laboratory indicated that quarterly audits will be performed and suggested that the audits results would be reviewed within the laboratory's QMPI Program. However, the laboratory did not establish the procedure by which these quarterly audits are to be conducted. In its submission, the laboratory indicates that a "tracer audit may [emphasis added] be used," but did not provided a protocol for a "tracer audit," the means by which a "tracer audit" would be documented, and whether the results of a "tracer audit" would be the information reviewed by the QMPI Program.

The laboratory failed to adequately address this deficiency and provide acceptable evidence of correction consisting of the required documentation and information set forth above and in our January 25, 2016 letter.

41