KETONES II
TABLE 1
MW: TABLE 1
METHOD: 1301, Issue 2
CAS: Table 1
RTECS: Table 1
EVALUATION: FULL COMPOUNDS (3) and (5): PARTIAL
OSHA/NIOSH/ACGIH: Table 1
COMPOUNDS: (Synonyms in Table 1)
1301
Issue 1: 15 February 1984 Issue 2: 15 August 1994
PROPERTIES: Table 1
(1) camphor (2) mesityl oxide
(3) 5-methyl-3-heptanone (4) methyl-(n-amyl)-ketone
SAMPLING
(5) ethyl butyl ketone
MEASUREMENT
SAMPLER:
SOLID SORBENT TUBE (coconut shell charcoal, 100 mg/50 mg)
FLOW RATE:
0.01 to 0.2 L/min
VOL-MIN: -MAX:
1L 25 L
SHIPMENT:
routine
TECHNIQUE:
GAS CHROMATOGRAPHY, FID
ANALYTE:
compounds above
DESORPTION:
1 mL 1% methanol in CS 2; stand in 30 min
INJECTION VOLUME: COLUMN: SAMPLE STABILITY:
unknown
FIELD BLANKS:
2 to 10 field blanks per set
5 µL
stainless steel (3 m x 3-mm ID), 10% FFAP on 80/100 Chromosorb W-AW
TEMPERATURE-INJECTOR: -DETECTOR: -COLUMN:
200 °C 300 °C 100 to 200 °C @ 10 °C/min
CARRIER GAS:
N 2 or He, 30 mL/min
CALIBRATION:
standard solutions of analyte in elution solvent
ACCURACY RANGE STUDIED:
see EVALUATION OF METHOD
BIAS:
see EVALUATION OF METHOD
ˆ rT): OVERALL PRECISION (S METHOD ACCURACY:
see EVALUATION OF
RANGE AND PRECISION:
see EVALUATION OF METHOD
ESTIMATED LOD: 0.05 mg per sample
see EVALUATION OF METHOD
APPLICABILITY: This method was developed to give better desorption than obtainable with carbon disulfide extraction of the charcoal. This method can be used in paint and resin manufacturing plants.
INTERFERENCES: None reported. Alternate columns, e.g., 4 m 10% SP-2100/0.1% Carbowax 1500 on Supelcoport 100/120 or DB-1 fused silica capillary may be used.
OTHER METHODS: This method combines and replaces Methods S10, S12, S13, S15 and S16 [1].
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/94