ALCOHOLS II
Table 1
MW: Table 1
METHOD: 1401, Issue 2
CAS: Table 2
Issue 1: 15 February 1984 Issue 2: 15 August 1994
PROPERTIES: Table 1
(1) (2) (3) (4)
n-butyl alcohol: 1-butanol; n-Butanol; Propyl carbinol sec-butyl alcohol: 2-butanol; methyl ethyl carbinol; 2-hydroxybutane. isobutyl alcohol: 2-methyl-1-propanol; isopropyl carbinol; IBA. n-propyl alcohol: 1-propanol; ethyl carbinol.
SAMPLING SAMPLER:
RTECS: Table 2
EVALUATION: FULL
OSHA : Table 2 NIOSH: Table 2 ACGIH: Table 2 COMPOUNDS and SYNONYMS:
1401
MEASUREMENT
SOLID SORBENT TUBE (coconut shell charcoal, 100 mg/50 mg)
TECHNIQUE:
GAS CHROMATOGRAPHY, FID
ANALYTE:
compounds above
DESORPTION:
1 mL 1% 2-propanol in CS2
INJECTION VOLUME:
5 µL
FLOW RATE: 0.01 to 0.2 L/min
VOL-MIN: -MAX:
(1) 2L 10 L
(2) 2L 10 L
(3) 2L 10 L
SHIPMENT:
routine
SAMPLE STABILITY:
unknown; store in freezer
BLANKS:
2 to 10 field blanks per set
(4) 1L 10 L
TEMPERATURE-INJECTION: -DETECTOR: -COLUMN:
200 °C 250-300 °C 75 °C
CARRIER GAS:
N2 or He, 20 mL/min
COLUMN:
glass, 3 m x 2-mm ID, 10% SP-1000 on 80/100 mesh Chromosorb WHP, or equivalent
CALIBRATION:
solutions of analyte in eluent (internal standard optional)
ACCURACY RANGE STUDIED:
see EVALUATION OF METHOD
BIAS:
not significant [1]
ˆ ): OVERALL PRECISION (S rT METHOD ACCURACY:
± 15%
see EVALUATION OF
RANGE AND PRECISION:
see EVALUATION OF METHOD
ESTIMATED LOD: 0.01 mg per sample [2]
APPLICABILITY: This method may be used to determine two or more analytes simultaneously by varying GC conditions (e.g., temperature programming).
INTERFERENCES: The methods were validated using a 3 m x 3-mm ID stainless steel column packed with 10% FFAP on Chromosorb W-AW; other columns with equal or better resolution (e.g., capillary) may be used.
OTHER METHODS: This method combines and replaces Methods S66, S53, S64 and S62 [3].
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/94