PHENYL ETHER
(C6H5)2O
MW: 170.22
METHOD: 1617, Issue 1
CAS: 101-84-8
1617
RTECS: KN8970000
EVALUATION: PARTIAL
OSHA : 1 ppm NIOSH: 1 ppm ACGIH: STEL 2 ppm (1 ppm = 6.96 mg/m 3)
PROPERTIES:
Issue 1: 15 August 1994
solid, geranium-like odor; MP 27 °C; BP -257 °C; d 1.072 g/mL @ 22 °C; VP 2.7 Pa (0.02 mm Hg) @ 25 °C; vapor density 5.86 (air = 1); flash pt. 115 °C; flammable limits in air 0.8 to 1.5%
SYNONYMS: diphenyl ether; diphenyl oxide; 1,1'-oxybisbenzene.
SAMPLING
MEASUREMENT
SAMPLER:
CHARCOAL TUBE (coconut shell charcoal, 100 mg/50 mg)
FLOW RATE:
0.01 to 0.2 L/min
VOL-MIN: -MAX:
1 L @ 1 ppm 50 L
TECHNIQUE:
GAS CHROMATOGRAPHY, FID
ANALYTE:
phenyl ether
EXTRACTION:
0.5 mL carbon disulfide, 30 min
INJECTION VOLUME:
5 µL
SHIPMENT: SAMPLE STABILITY:
not determined
TEMPERATURE-INJECTION: -DETECTOR: -COLUMN:
230 °C 265 °C 215 °C
BLANKS:
2 to 10 field blanks per set
COLUMN:
10 ft x 1/8-in. ID stainless steel column packed with 10% FFAP on 80/100 mesh, acid washed DMCS Chromosorb W
DETECTOR:
flame ionization detector
CALIBRATION:
standard solutions of phenyl ether in hexane
RANGE:
7 to 210 µg per sample [1,2]
ACCURACY RANGE STUDIED:
3.0 to 13.0 mg/m 3 [1] (10-L samples)
BIAS:
0.06%
ˆ ): OVERALL PRECISION (S rT
0.07 [1,2]
ACCURACY:
± 13.7%
ESTIMATED LOD: 0.7 µg per sample [1,2] PRECISION (Sr):
0.048 [1,2]
APPLICABILITY: The working range is 0.1 to 9 ppm (0.7 to 60 mg/m 3) for a 10-L air sample. Better sensitivity may be achieved with a capillary column with appropriate adjustment in the instrumental conditions. A DB-wax (polyethelyene glycol) or e quivalent is similar to the FFAP packed column.
INTERFERENCES: None identified.
OTHER METHODS: This revises Method S72 [3].
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/94