CARBON DIOXIDE 6603
CO₂ MW: 44.01 CAS: 124-38-9 RTECS: FF6400000

METHOD: 6603 EVALUATION: FULL Issue 1: 15 May 1989
Issue 2: 15 August 1994

OSHA: 5000 ppm; STEL 30000 ppm PROPERTIES: gas; sublimes @ -78.5 °C
NIOSH: 5000 ppm; STEL 30000 ppm
ACGIH: 5000 ppm; STEL 30000 ppm
(1ppm = 1.8 mg/m³ @ NTP)

SYNONYMS: carbonic acid; dry ice


SAMPLING


SAMPLER: GAS SAMPLING BAG
FLOW RATE: 0.02 to 0.1 L/min; fill bag ≤ 80% of capacity; spot samples (step 2.a.) or TWA samples (step 2.b.)
VOL-MIN:
  -MAX:
SHIPMENT:
SAMPLE STABILITY: at least 7 days @ 25 °C [1]
BLANKS: in a bag from a non-work area

ACCURACY


RANGE STUDIED: 2270 - 10000 ppm [3] (3.5-L samples)
BIAS: -2.5% [3]
OVERALL PRECISION (ŜrT): 0.014 [1]
ACCURACY: ±5.3% [1]

MEASUREMENT


TECHNIQUE: GAS CHROMATOGRAPHY (PORTABLE),
TCD
ANALYTE: carbon dioxide
DESORPTION:
INJECTION VOLUME:
TEMPERATURE-INJECTION: ambient
     -DETECTOR: 70 °C
      -COLUMN: ambient
CARRIER GAS: He, 100 mL/min
COLUMN: 1.5 m x 6-mm ID stainless steel, packed with 80/100 mesh Poropak QS
CALIBRATION: bag standards or calibrated gas mixtures
RANGE: 500 - 15000 ppm [2]
ESTIMATED LOD: 1 ppm
PRECISION (S̄r): 0.005 [1]

APPLICABILITY: The working range is 500 to 15000 ppm (900 to 27000 mg/m³) in relatively non-complex atmospheres.


INTERFERENCES: Any compound having the same or nearly the same retention time as carbon dioxide on the column in use.


OTHER METHODS: This is method S249 in revised format [2].