Page:NIOSH Manual of Analytical Methods - 5526.pdf/5

From Wikisource
Jump to navigation Jump to search
This page needs to be proofread.

METHYLTIN CHLORIDES: METHOD 5526, Issue 1, dated 15 March 2003 - Page 5 of 5 TABLE 2. METHOD EVALUATION [1] Range Studied

Bias (%)

Precision(S rT) (%)

Accuracy (%)

Trimethyltin chloride

0.46 - 9.29 :g/sample (tin)

0.3

5.2

11

Dim eth yltin dichloride

0.42 - 9.17 :g/sample (tin)

-2.5

5.8

12

Methyltin chloride

0.07 - 10.48 :g/sample (tin)

-5.2

7.6

16

Figure 1:

Sample Chromatogram

NIOSH Manual of Analytical Methods (NMAM), Fourth Edition